
1
1451 Belcher Road South, Largo, FL 33773 • USA
• Tel +1 (727) 447-6140 •
Fax (727) 442-5699 •
[email protected]F-5100 Insertion Thermal Mass Flow Meter Manual 05/15 - 0663-20 / 18333 Page A-4
Method 3 (Ambient Pressure Air Calibration Test)
1. While rmly grasping the enclosure, loosen the compression
tting at the top of the ball valve.
2. Slowly withdraw the stem until the ats located on the sensor
are exposed.
3. Carefully close the ball valve and fully withdraw the ow
sensor.
4. Place the ow sensor in to an empty water bottle.
5. Wait up to 2 – 3 minutes to allow the sensor reading to stabilize at zero and note
the mW reading on the display.
6. Compare the reading to the zero ow air calibration shown on the laminated tag
attached to the sensor. The two readings should agree within ± 5mW.
7. At the conclusion of the test, re-install the sensor as per the instruction manual.
IMPORTANT NOTE
The ambient pressure air calibration test can be used to validate the sensor calibration independent
of the process uid. As such, this test can be a useful diagnostic tool.
Method 2 (Used when it is not possible to stop ow in the pipe.)
1. While rmly grasping the enclosure, loosen the
compression tting at the top of the ball valve.
2. Slowly withdraw the stem until the bottom of the ats
located on the sensor is just exposed.
3. Re-tighten the compression tting to ½ turn beyond
hand tight.
4. Carefully close the ball valve.
5. Wait up to 2 – 3 minutes to allow the sensor reading to
stabilize at zero and note the mW reading on the display.
6. Compare the reading to that stamped on the nameplate.
The two readings should agree within ± 5mW.
7. At the conclusion of the test, re-install the sensor as per
the instruction manual.
Interpreting Test Results
Zero ow calibration test readings that are within ± 5mW of the original calibration value indicate
that the sensor calibration is still valid.
Zero ow calibration test readings that are within ± 10mW of the original calibration value indicate
that the sensor calibration is still valid, but the test results were likely inuenced by external factors.
These include any or all of the following:
1. Sensor orientation (installed horizontally).
2. Dirt or debris is coating all or part of the sensor assembly.
3. Process uid composition different from the uid used during calibration.
(Use ambient air calibration test to conrm.)
4. Residual ow readings during test due to leaking valves or sensor movement.
(Repeat test using alternate method.)
Komentarze do niniejszej Instrukcji